ICU VENTILATOR TT24

The ventilator makes a good performance in operation room, ICU department and emergency treatment. It used to assist or replace the spontaneous breathing for adult pediatric and neonatal more than 2kg. Due to the flexible configuration, good quality and competitive price.

FEATURES:
  • 10.4” TFT screen displays the ventilation parameters, alarm information, and oscillograms, make every operation more easily.
  • Multiple ventilation mode can meet different clinical requirements ( IPPV, A/C,SIMV,SPONT/CPAP,SIGH,MANUAL).
  • Multiple parameters monitoring interface,make every parameter clear for users to know patient condition in all respects.
  • Pressure-Time,Flow-Time oscillograms and high precision oxygen concentration detection function included.
  • Humidifier can heat and wet breathing gas, makes it comfortable for patient to breathe.
  • Rapid oxygen supply, automatically offer high flow rate oxygen within two minutes.
  • High temperature resistance breathing circuit is reusable and anti-pollution.
  • 14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting.
  • Easy to move with four casters, easy to stop with two brakes.
  • Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
  • Compact long life internal battery can provide emergency power, avoid risk of patient.
  • Self-check before operation,eliminate system mistake.

Specifications

GENERAL PARAMETERS

Ventilation modes: IPPV, A/C, SIMV, SPONT/CPAP, MANUAL

VENTILATOR PARAMETER RANGES

Tidal volume(Vt): 0.20 ~ 1500 mL

Frequency (Freq): 1bpm ~ 100 bpm

Oxygen concentration: 45%-100%

I/E: 4:1~1:8

PEEP: 0 cmH2O ~ 20 cmH2O

SIGH: 0 (off) 1 / 100 ~ 5 / 100Pressure 

Pressure Triggering sensitivity (Ptr) -20 cmH2O ~ 0 cmH2O (Based on PEEP)

Apnea Ventilation:  OFF, 5 s ~ 60 s

MONITORED PARAMETERS

Frequency (Freq)0 /min ~ 100 /min

Tidal volume(Vt)0 mL ~ 2000 mL

MV: 0 L/min ~ 99 L/min

Airway pressure: 0 cmH2O ~ 100 cmH2O

Oxygen concentration: 15 % ~ 100 %

ALARM AND PROTECTION

The AC power failure alarm: Power failure or no connection

Internal battery backup low voltage alarm: ≤11.3±0.3V

No tidal volume: No tidal volume within 6s

High Minute Volume alarm: 1L/min-99L/min

Low Minute Volume alarm: 1L/min-20L/min

High Airway pressure alarm: 20cmH2O-100cmH2O

Low Airway pressure alarm: 0cmH2O-20cmH2O

High oxygen concentration alarm: 19%-100%

Low oxygen concentration alarm: 18%-99%

Continuous pressure alarm: (PEEP+1.5 kPa) over 16s

Suffocation warning: 5-60s no spontaneous ventilation

WORKING CONDITIONS

Gas source: O2

Pressure: 280kPa-600kPa

Voltage: ~220V ±22V

Power frequency: 50Hz ± 1Hz

Input power: 40 VA

DIMENSIONS

Main Components: 560 x 560 x 610 mm

G.W.: 40 Kg, N.W.: 17 Kg

Trolley: 650 x 690 x 520 mm

G.W.: 43 Kg, N.W.: 18 Kg

 

Certification: CE 0197, ISO 13485, CFDA, FSC

 

 